1142 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:USBPd:TRIGger:HEADer:CMESsage
(see page 1610)
Command Syntax
:SBUS<n>:USBPd:TRIGger:HEADer:CMESsage <type>
<type> ::= {GOODcrc | GOTOmin | ACCept | REJect | PING | PSRDy
| GSRCap | GSNCap | DRSWap | PRSWap | VCSWap | WAIT | SRST | GSCX
| GSTatus | FRSWap | GPSTatus | GCCodes}
When the Header Content trigger mode is selected (with the
":SBUS<n>:USBPd:TRIGger HEADer" command) and the Control Message header
type is selected (with the ":SBUS<n>:USBPd:TRIGger:HEADer CMESsage"
command), the :SBUS<n>:USBPd:TRIGger:HEADer:CMESsage command selects
the control message type:
GOODcrc — GoodCRC
•GOTOmin GotoMin
ACCept — Accept
•REJect Reject
•PING Ping
•PSRDy PS_RDY
GSRCap — Get_Source_Cap
GSNCap — Get_Sink_Cap
•DRSWap DR_Swap
PRSWap — PR_Swap
VCSWap — VCONN_Swap
•WAIT Wait
•SRST Soft_Reset
GSCX — Get_Source_Cap_Extended
GSTatus — Get_Status
FRSWap — FR_Swap
GPSTatus — Get_PPS_Status
GCCodes — Get_Country_Codes
Query Syntax
:SBUS<n>:USBPd:TRIGger:HEADer:CMESsage?
The :SBUS<n>:USBPd:TRIGger:HEADer:CMESsage? query returns the selected
control message type.
Return Format
<type><NL>
<type> ::= {GOOD | GOTO | ACC | REJ | PING | PSRD | GSRC | GSNC | DRSW
| PRSW | VCSW | WAIT | SRST | GSCX | GST | FRSW | GPST | GCC}