1272 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33 :SYSTem Commands
:SYSTem:RLOGger:DISPlay
(see page 1610)
Command Syntax
:SYSTem:RLOGger:DISPlay {{0 | OFF} | {1 | ON}}
The :SYSTem:RLOGger:DISPlay command enables or disables the screen display
of logged remote commands and their return values (if applicable).
Query Syntax
:SYSTem:RLOGger:DISPlay?
The :SYSTem:RLOGger:DISPlay? query returns whether the screen display for
remote command logging is enabled or disabled.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also ":SYSTem:RLOGger" on page 1270
":SYSTem:RLOGger:DESTination" on page 1271
":SYSTem:RLOGger:FNAMe" on page 1273
":SYSTem:RLOGger:STATe" on page 1274
":SYSTem:RLOGger:TRANsparent" on page 1275
":SYSTem:RLOGger:WMODe" on page 1276