874 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:A429:TRIGger:RANGe
(see page 1610)
Command Syntax
:SBUS<n>:A429:TRIGger:RANGe <min>,<max>
<min> ::= 8-bit integer in decimal, <hex>, <octal>, or <string>
from 0-255
<max> ::= 8-bit integer in decimal, <hex>, <octal>, or <string>
from 0-255
<hex> ::= #Hnn where n ::= {0,..,9 | A,..,F}
<octal> ::= #Qnnn where n ::= {0,..,7}
<string> ::= "0xnn" where n::= {0,..,9 | A,..,F}
The :SBUS<n>:A429:TRIGger:RANGe command defines a range of ARINC 429
label values. This range is used when the LRANge trigger type is selected.
Query Syntax
:SBUS<n>:A429:TRIGger:RANGe?
The :SBUS<n>:A429:TRIGger:RANGe? query returns the current label values in
decimal format.
Return Format
<min>,<max><NL> in decimal format
Errors • "-241, Hardware missing" on page 1551
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":SBUS<n>:A429:TRIGger:TYPE" on page 875

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals