370 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
12 :COMPliance Commands
:COMPliance:USB:RUN
(see page 1610)
Command Syntax
:COMPliance:USB:RUN
The :COMPliance:USB:RUN command runs the selected signal quality test.
Please be patient as tests can take several minutes to complete.
When tests are run, the oscilloscope stops acquisitions if they are running,
analyzes the data on screen, and then displays the results. The analyzed
acquisition remains on screen and can be viewed by moving or dismissing the
results dialog.
You can run tests when the oscilloscope is already stopped. This is useful in the
case of embedded hosts (and other cases) where the device under test (DUT) may
not easily be placed in a test mode, and you want to analyze already acquired
data.
See Also • ":COMPliance:USB:TEST" on page 375
• ":COMPliance:USB:AUTosetup" on page 368
• "*OPC (Operation Complete)" on page 240
• ":SAVE:COMPliance:USB[:STARt]" on page 826