:MEASure Power Commands 25
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 647
:MEASure:APParent
(see page 1610)
Command Syntax
:MEASure:APParent [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:APParent command installs an apparent power measurement on
screen.
The <source1> parameter is the channel probing voltage and the <source2>
parameter is the channel probing current. These sources can also be specified by
the :MEASure:SOURce command.
Apparent power is a measure of power quality. It is the portion of AC line power
flow due to stored energy which returns to the source in each cycle.
IRMS * VRMS
Query Syntax
:MEASure:APParent? [<source1>][,<source2>]
The :MEASure:APParent query returns the measured apparent power.
Return Format
<return_value><NL>
<return_value> ::= the apparent power value in NR3 format
See Also • ":MEASure:SOURce" on page 615
• ":POWer:QUALity:APPLy" on page 774

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals