1316 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:DELay:TRIGger:SLOPe
(see page 1610)
Command Syntax
:TRIGger:DELay:TRIGger:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :TRIGger:DELay:TRIGger:SLOPe command specifies rising (POSitive) or falling
(NEGative) for the trigger edge in the Edge Then Edge trigger.
Query Syntax
:TRIGger:DELay:TRIGger:SLOPe?
The :TRIGger:DELay:TRIGger:SLOPe? query returns the current trigger edge slope
setting.
Return Format
<slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 1295
":TRIGger:DELay:TRIGger:SOURce" on page 1317
":TRIGger:DELay:TDELay:TIME" on page 1314
":TRIGger:DELay:TRIGger:COUNt" on page 1315