452 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
19 :FRANalysis Commands
:FRANalysis:WGEN:LOAD
(see page 1610)
Command Syntax
:FRANalysis:WGEN:LOAD <impedance>
<impedance> ::= {ONEMeg | FIFTy}
The :FRANalysis:WGEN:LOAD command selects the expected output load
impedance.
The output impedance of the Gen Out BNC is fixed at 50 ohms. However, the
output load selection lets the waveform generator display the correct amplitude
and offset levels for the expected output load.
If the actual load impedance is different than the selected value, the displayed
amplitude and offset levels will be incorrect.
Query Syntax
:FRANalysis:WGEN:LOAD?
The :FRANalysis:WGEN:LOAD? query returns the current expected output load
impedance.
Return Format
<impedance><NL>
<impedance> ::= {ONEM | FIFT}
See Also • ":FRANalysis:DATA" on page 441
• ":FRANalysis:ENABle" on page 442
• ":FRANalysis:FREQuency:STARt" on page 445
• ":FRANalysis:FREQuency:STOP" on page 446
• ":FRANalysis:RUN" on page 448
• ":FRANalysis:SOURce:INPut" on page 449
• ":FRANalysis:SOURce:OUTPut" on page 450
• ":FRANalysis:WGEN:VOLTage" on page 453