:FRANalysis Commands 19
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 441
:FRANalysis:DATA
(see page 1610)
Query Syntax
:FRANalysis:DATA? [SWEep | SINGle]
The :FRANalysis:DATA? query returns the frequency response analysis data.
The data is returned in four comma-separated columns of data for each step in the
sweep: Frequency (Hz), Amplitude (Vpp), Gain (dB), and Phase (°).
You can use the :FRANalysis:TRACe command to specify whether to include gain,
phase, both gain and phase, or neither in the frequency response analysis results.
The SWEep or SINGle option specifies whether to get the data from a sweep or
single-frequency analysis (see :FRANalysis:FREQuency:MODE). If this option is not
specified, the data from the sweep analysis is returned by default.
Return Format
<binary_block><NL>
<binary_block> ::= comma-separated data with newlines at the end of each
row
See Also • ":FRANalysis:ENABle" on page 442
• ":FRANalysis:FREQuency:MODE" on page 443
• ":FRANalysis:FREQuency:SINGle" on page 444
• ":FRANalysis:FREQuency:STARt" on page 445
• ":FRANalysis:FREQuency:STOP" on page 446
• ":FRANalysis:RUN" on page 448
• ":FRANalysis:SOURce:INPut" on page 449
• ":FRANalysis:SOURce:OUTPut" on page 450
• ":FRANalysis:TRACe" on page 451
• ":FRANalysis:WGEN:LOAD" on page 452
• ":FRANalysis:WGEN:VOLTage" on page 453