440 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
19 :FRANalysis Commands
Introduction to
:FRANalysis
Commands
The FRANalysis subsystem controls the Frequency Response Analysis feature in
the oscilloscope.
The Frequency Response Analysis (FRA) feature controls the built-in waveform
generator to sweep a sine wave across a range of frequencies while measuring the
input to and output from a device under test (DUT). At each frequency, gain (A)
and phase are measured and plotted on a frequency response chart.
Reporting the Setup
Use :FRANalysis? to query setup information for the FRANalysis subsystem.
Return Format
The following is a sample response from the :FRANalysis? query. In this case, the
query was issued following a *RST command.
:FRAN:SOUR:INP CHAN1;OUTP CHAN2;:FRAN:FREQ:STAR +100E+00;
STOP +20.000000E+06;:FRAN:WGEN:VOLT +200.0E-03;LOAD FIFT
:FRANalysis:SOURce:IN
Put <source> (see
page 449)
:FRANalysis:SOURce:IN
Put? (see page 449)
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:FRANalysis:SOURce:OU
TPut <source> (see
page 450)
:FRANalysis:SOURce:OU
TPut? (see page 450)
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:FRANalysis:TRACe
<selection> (see
page 451)
:FRANalysis:TRACe?
(see page 451)
<selection> ::= {NONE | ALL |
GAIN | PHASe}[,{GAIN | PHASe}]
:FRANalysis:WGEN:LOAD
<impedance> (see
page 452)
:FRANalysis:WGEN:LOAD
? (see page 452)
<impedance> ::= {ONEMeg | FIFTy}
:FRANalysis:WGEN:VOLT
age
<amplitude>,[<range>]
(see page 453)
:FRANalysis:WGEN:VOLT
age? [<range>] (see
page 453)
<amplitude> ::= amplitude in
volts in NR3 format
<range> ::= {F20HZ | F100HZ |
F1KHZ | F10KHZ | F100KHZ | F1MHZ
| F10MHZ | F20MHZ}
:FRANalysis:WGEN:VOLT
age:PROFile {{0 |
OFF} | {1 | ON}} (see
page 454)
:FRANalysis:WGEN:VOLT
age:PROFile? (see
page 454)
{0 | 1}
Table 105 :FRANalysis Commands Summary (continued)
Command Query Options and Query Returns

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals