EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1163 background imageLoading...
Page #1163 background image
:SEARch Commands 32
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1163
:SEARch:GLITch:RANGe
(see page 1610)
Command Syntax
:SEARch:GLITch:RANGe <less_than_time>[suffix],
<greater_than_time>[suffix]
<less_than_time> ::= (15 ns - 10 seconds) in NR3 format
<greater_than_time> ::= (10 ns - 9.99 seconds) in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :SEARch:GLITch:RANGe command sets the pulse width duration for the
selected :SEARch:GLITch:SOURce. You can enter the parameters in any order —
the smaller value becomes the <greater_than_time> and the larger value becomes
the <less_than_time>.
Query Syntax
:SEARch:GLITch:RANGe?
The :SEARch:GLITch:RANGe? query returns the pulse width duration time for
:SEARch:GLITch:SOURce.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also Chapter 32, “:SEARch Commands,” starting on page 1149
":SEARch:GLITch:SOURce" on page 1164
":SEARch:GLITch:QUALifier" on page 1162
":SEARch:MODE" on page 1153

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals