1288 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
34 :TIMebase Commands
:TIMebase:REFerence:LOCation
(see page 1610)
Command Syntax
:TIMebase:REFerence:LOCation <loc>
<loc> ::= 0.0 to 1.0 in NR3 format
When the :TIMebase:REFerence is set to CUSTom, the
:TIMebase:REFerence:LOCation command lets you place the time reference
location at a percent of the graticule width (where 0.0 is the left edge and 1.0 is
the right edge).
Query Syntax
:TIMebase:REFerence:LOCation?
The :TIMebase:REFerence:LOCation? query returns the time base reference
custom location setting.
Return Format
<loc><NL>
<loc> ::= 0.0 to 1.0 in NR3 format
See Also • ":TIMebase:REFerence" on page 1287