:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1029
:SBUS<n>:NRZ:LOGic
(see page 1610)
Command Syntax
:SBUS<n>:NRZ:LOGic <logic>
<logic> ::= {HIGH | LOW}
The :SBUS<n>:NRZ:LOGic command specifies the polarity of the NRZ signal:
HIGH — specifies that a positive voltage is used to encode a bit value of logic 1
(and a negative voltage encodes a bit value of logic 0).
LOW — specifies that a negative voltage is used to encode a bit value of logic 1.
Query Syntax
:SBUS<n>:NRZ:LOGic?
The :SBUS<n>:NRZ:LOGic? query returns the polarity setting.
Return Format
<logic><NL>
<logic> ::= {HIGH | LOW}
See Also ":SBUS<n>:NRZ:BASE" on page 1020