:CHANnel<n> Commands 11
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 365
:CHANnel<n>:UNITs
(see page 1610)
Command Syntax
:CHANnel<n>:UNITs <units>
<units> ::= {VOLT | AMPere}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:UNITs command sets the measurement units for the connected
probe. Select VOLT for a voltage probe and select AMPere for a current probe.
Measurement results, channel sensitivity, and trigger level will reflect the
measurement units you select.
Query Syntax
:CHANnel<n>:UNITs?
The :CHANnel<n>:UNITs? query returns the current units setting for the specified
channel.
Return Format
<units><NL>
<units> ::= {VOLT | AMP}
See Also • "Introduction to :CHANnel<n> Commands" on page 345
• ":CHANnel<n>:RANGe" on page 363
• ":CHANnel<n>:PROBe" on page 354
• ":EXTernal:UNITs" on page 437