EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Root (:) Commands 7
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 263
:AUToscale:AMODE
(see page 1610)
Command Syntax
:AUToscale:AMODE <value>
<value> ::= {NORMal | CURRent}
The :AUToscale:AMODE command specifies the acquisition mode that is set by
subsequent :AUToscales.
When NORMal is selected, an :AUToscale command sets the NORMal
acquisition type and the RTIMe (real-time) acquisition mode.
When CURRent is selected, the current acquisition type and mode are kept on
subsequent :AUToscales.
Use the :ACQuire:TYPE and :ACQuire:MODE commands to set the acquisition type
and mode.
Query Syntax
:AUToscale:AMODE?
The :AUToscale:AMODE? query returns the autoscale acquire mode setting.
Return Format
<value><NL>
<value> ::= {NORM | CURR}
See Also "Introduction to Root (:) Commands" on page 258
":AUToscale" on page 261
":AUToscale:CHANnels" on page 264
":ACQuire:TYPE" on page 318
":ACQuire:MODE" on page 307

Table of Contents

Related product manuals