:FUNCtion<m> Commands 20
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 501
:FUNCtion<m>:TRENd:NMEasurement
(see page 1610)
Command Syntax
:FUNCtion<m>:TRENd:NMEasurement MEAS<n>
<n> ::= # of installed measurement, from 1 to 10
<m> ::= 1 to (# math functions) in NR1 format
The :FUNCtion<m>:TRENd:NMEasurement command selects the measurement
whose trend is shown in the math waveform.
There are 10 locations (or slots) that installed measurements can occupy. The
MEAS<n> parameter specifies the location of the measurement whose trend you
want to analyze.
You can view the trend math function waveform for these installed measurements:
• :MEASure:VAVerage, <interval> ::= CYCLe
• :MEASure:VRMS, <type> ::= AC (AC RMS)
• :MEASure:VRATio, <interval> ::= CYCLe
• :MEASure:PERiod
• :MEASure:FREQuency
• :MEASure:PWIDth
• :MEASure:NWIDth
• :MEASure:DUTYcycle
• :MEASure:NDUTy
• :MEASure:RISetime
• :MEASure:FALLtime
Query Syntax
:FUNCtion<m>:TRENd:NMEasurement?
The :FUNCtion<m>:TRENd:NMEasurement? query returns the selected
measurement.
If no measurements are installed, the :FUNCtion<m>:TRENd:NMEasurement?
query will return NONE.
Return Format
MEAS<n><NL>
<n> ::= # of installed measurement, from 1 to 10
See Also • ":FUNCtion<m>:OPERation" on page 490

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals