:MEASure Commands 24
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 553
:MEASure:DUAL:VRMS
[<interval>]
[,<type>]
[,<source1>][,<source
2>] (see page 588)
:MEASure:DUAL:VRMS?
[<interval>]
[,<type>]
[,<source1>][,<source
2>] (see page 588)
<interval> ::= {CYCLe | DISPlay}
<type> ::= {AC | DC}
<source1>,<source2> ::=
CHANnel<n> with N2820A probe
connected
<n> ::= 1 to (# analog channels)
in NR1 format
<return_value> ::= calculated
RMS voltage in NR3 format
:MEASure:DUTYcycle
[<source>] (see
page 589)
:MEASure:DUTYcycle?
[<source>] (see
page 589)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> |
WMEMory<r>} for DSO models
<source> ::= {CHANnel<n> |
DIGital<d> | FUNCtion<m> |
MATH<m> | WMEMory<r>} for MSO
models
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format
<return_value> ::= ratio of
positive pulse width to period in
NR3 format
Table 110 :MEASure Commands Summary (continued)
Command Query Options and Query Returns

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals