234 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
6 Common (*) Commands
*ESE (Standard Event Status Enable)
(see page 1610)
Command Syntax
*ESE <mask_argument>
<mask_argument> ::= integer from 0 to 255
The *ESE common command sets the bits in the Standard Event Status Enable
Register. The Standard Event Status Enable Register contains a mask value for the
bits to be enabled in the Standard Event Status Register. A "1" in the Standard
Event Status Enable Register enables the corresponding bit in the Standard Event
Status Register. A zero disables the bit.
Query Syntax
*ESE?
Table 79 Standard Event Status Enable (ESE)
Bit Name Description When Set (1 = High = True), Enables:
7 PON Power On Event when an OFF to ON transition occurs.
6 URQ User Request Event when a front-panel key is pressed.
5 CME Command Error Event when a command error is detected.
4 EXE Execution Error Event when an execution error is detected.
3 DDE Device Dependent Error Event when a device-dependent error is detected.
2 QYE Query Error Event when a query error is detected.
1 RQL Request Control Event when the device is requesting control. (Not
used.)
0 OPC Operation Complete Event when an operation is complete.
To ESB bit in
Statu s Byte
Register
Standard Event
Status Enable
(Mask) Register
Standard Event
Status Register
*ESR?
*ESE
*ESE?
PON URQ EXE DDE QYE RQL OPCCM E
OR
10234567
128 64 16 8 4 2 132