Page #276 background imageLoading...
Page #276 background image
276 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
7 Root (:) Commands
:MTERegister[:EVENt] (Mask Test Event Event Register)
(see page 1610)
Query Syntax
:MTERegister[:EVENt]?
The :MTERegister[:EVENt]? query returns the integer value contained in the Mask
Test Event Event Register and clears the register.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also • "Introduction to Root (:) Commands" on page 258
• ":CHANnel<n>:PROTection" on page 362
• ":OPEE (Operation Status Enable Register)" on page 278
• ":OPERegister:CONDition (Operation Status Condition Register)" on page 280

To MTE bit in
Operation Status
Condition Register
OR
:MTEenable
:MTEenable?
Mask Test Event Enable (Mask) Register
1023456798101112131415
:MTERegister[:EVENt]?
Mask Test Event Event Register
Fail
Com-
plete
Auto
Mask
Started
Pass
Table 88 Mask Test Event Event Register
Bit Name Description When Set (1 = High = True), Indicates:
15-11 --- --- (Not used.)
10 Auto
Mask
Auto Mask Created Auto mask creation completed.
9 Pass Mask Test Pass The mask test passed.
8 Started Mask Testing Started Mask testing started.
7-2 --- --- (Not used.)
1 Fail Mask Test Fail The mask test failed.
0 Complet
e
Mask Test Complete The mask test is complete.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals