:ACQuire Commands 8
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 301
averaged together to provide the value for one display point. The slower the sweep
speed, the greater the number of samples that are averaged together for each
display point.
Peak Detect
The :ACQuire:TYPE PEAK command sets the oscilloscope in the peak detect mode.
In this mode, :ACQuire:COUNt has no meaning.
Reporting the Setup
Use :ACQuire? to query setup information for the ACQuire subsystem.
Return Format
The following is a sample response from the :ACQuire? query. In this case, the
query was issued following a *RST command.
:ACQ:MODE RTIM;TYPE NORM;COMP 100;COUNT 8;SEGM:COUN 2