:MEASure Commands 24
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 581
In the :MEASure:DEFine command, you can set upper, middle, and lower threshold
values. It is the middle threshold value that is used when performing the delay
query. The standard upper, middle, and lower measurement thresholds are 90%,
50%, and 10% values between Vbase and Vtop. If you want to move the delay
measurement point nearer to Vtop or Vbase, you must change the threshold
values with the :MEASure:DEFine THResholds command.
Return Format
<value><NL>
<value> ::= floating-point number delay time in seconds in NR3 format
See Also • "Introduction to :MEASure Commands" on page 568
• ":MEASure:DELay:DEFine" on page 582
• ":MEASure:DEFine" on page 577
• ":MEASure:PHASe" on page 604