Commands Quick Reference 5
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 151
:MTESt:RMODe:SIGMa
<level> (see page 689)
:MTESt:RMODe:SIGMa?
(see page 689)
<level> ::= from 0.1 to 9.3 in
NR3 format
:MTESt:RMODe:TIME
<seconds> (see
page 690)
:MTESt:RMODe:TIME?
(see page 690)
<seconds> ::= from 1 to 86400 in
NR3 format
:MTESt:RMODe:WAVeform
s <count> (see
page 691)
:MTESt:RMODe:WAVeform
s? (see page 691)
<count> ::= number of waveforms
in NR1 format
:MTESt:SCALe:BIND {{0
| OFF} | {1 | ON}}
(see page 692)
:MTESt:SCALe:BIND?
(see page 692)
{0 | 1}
:MTESt:SCALe:X1
<x1_value> (see
page 693)
:MTESt:SCALe:X1? (see
page 693)
<x1_value> ::= X1 value in NR3
format
:MTESt:SCALe:XDELta
<xdelta_value> (see
page 694)
:MTESt:SCALe:XDELta?
(see page 694)
<xdelta_value> ::= X delta value
in NR3 format
:MTESt:SCALe:Y1
<y1_value> (see
page 695)
:MTESt:SCALe:Y1? (see
page 695)
<y1_value> ::= Y1 value in NR3
format
:MTESt:SCALe:Y2
<y2_value> (see
page 696)
:MTESt:SCALe:Y2? (see
page 696)
<y2_value> ::= Y2 value in NR3
format
:MTESt:SOURce
<source> (see
page 697)
:MTESt:SOURce? (see
page 697)
<source> ::= {CHANnel<n> | NONE}
<n> ::= {1 | 2 | 3 | 4} for 4ch
models
<n> ::= {1 | 2} for 2ch models
n/a :MTESt:TITLe? (see
page 698)
<title> ::= a string of up to 128
ASCII characters
Table 22 :MTESt Commands Summary (continued)
Command Query Options and Query Returns
Table 23 :POD<n> Commands Summary
Command Query Options and Query Returns
:POD<n>:DISPlay {{0 |
OFF} | {1 | ON}} (see
page 701)
:POD<n>:DISPlay? (see
page 701)
{0 | 1}
<n> ::= 1-2 in NR1 format

Table of Contents