EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Status Reporting 41
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1587
IO_COMPLETE = 1 << IOC_BIT # 1 leftshift 13 = 8192
# Test for armed.
STATUS_REGISTER = int(KsInfiniiVisionX.query(":OPER?"))
IO_COMPLETE_STATUS = STATUS_REGISTER & IOC_MASK
# Wait indefinitely until armed.
while IO_COMPLETE_STATUS != IO_COMPLETE:
# Check the status again after small delay.
time.sleep(0.1) # 100 ms delay to prevent excessive queries.
STATUS_REGISTER = int(KsInfiniiVisionX.query(":OPER?"))
IO_COMPLETE_STATUS = STATUS_REGISTER & IOC_MASK
time_in_seconds = str(time.clock() - now)
print "IO Operation Complete (from :OPER? query)."
print "It took %s seconds for IO operation to complete.\n" % \
time_in_seconds
# ====================================================================
# Main
# ====================================================================
# Connect and initialize oscilloscope
# --------------------------------------------------------------------
# Define VISA Resource Manager & Install directory
rm = visa.ResourceManager('C:\\Windows\\System32\\agvisa32.dll')
# Define and open the oscilloscope using the VISA address
KsInfiniiVisionX = rm.open_resource(VISA_ADDRESS)
# Set the Global Timeout
KsInfiniiVisionX.timeout = GLOBAL_TOUT
# Clear the instrument bus
KsInfiniiVisionX.clear()
# Reset the oscilloscope.
# KsInfiniiVisionX.write("*RST")
# Or comment out to use the current oscilloscope setup.
# Autoscale to set up vertical scale and trigger level on channels 1 and
2.
KsInfiniiVisionX.write(":CHANnel1:DISPlay ON")
KsInfiniiVisionX.write(":CHANnel2:DISPlay ON")
KsInfiniiVisionX.write(":AUToscale:CHANnels DISPlayed")
KsInfiniiVisionX.write(":AUToscale")
# Between :WMEMory<r>:SAVE commands, wait for IO complete.
# --------------------------------------------------------------------
# Clear all status registers before checking for new events.
KsInfiniiVisionX.write("*CLS")
KsInfiniiVisionX.write(":WMEMory1:SAVE CHANnel1")
wait_io_operation()
KsInfiniiVisionX.write(":WMEMory2:SAVE CHANnel2")
wait_io_operation()