Commands Quick Reference 5
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 161
:RECall:SETup[:STARt]
[<file_spec>] (see
page 818)
n/a <file_spec> ::= {<internal_loc>
| <file_name>}
<internal_loc> ::= 0-9; an
integer in NR1 format
<file_name> ::= quoted ASCII
string
:RECall:WMEMory<r>[:S
TARt] [<file_name> |
<data>] (see page 819)
n/a <r> ::= 1 to (# ref waveforms) in
NR1 format
<file_name> ::= quoted ASCII
string
If extension included in file
name, it must be ".h5".
<data> ::= binary block data in
IEEE 488.2 # format
Table 25 :RECall Commands Summary (continued)
Command Query Options and Query Returns
Table 26 :SAVE Commands Summary
Command Query Options and Query Returns
:SAVE:ARBitrary:[STAR
t] [<file_spec>][,
<wavegen_id>] (see
page 825)
n/a <file_spec> ::= {<internal_loc>
| <file_name>}
<internal_loc> ::= 0-3; an
integer in NR1 format
<file_name> ::= quoted ASCII
string
<wavegen_id> ::= {WGEN1 | WGEN2}
:SAVE:COMPliance:USB[
:STARt] [<file_name>]
(see page 826)
n/a <file_name> ::= quoted ASCII
string
:SAVE:FILename
<base_name> (see
page 827)
:SAVE:FILename? (see
page 827)
<base_name> ::= quoted ASCII
string
:SAVE:IMAGe[:STARt]
[<file_name>] (see
page 828)
n/a <file_name> ::= quoted ASCII
string
:SAVE:IMAGe:FACTors
{{0 | OFF} | {1 |
ON}} (see page 829)
:SAVE:IMAGe:FACTors?
(see page 829)
{0 | 1}
:SAVE:IMAGe:FORMat
<format> (see
page 830)
:SAVE:IMAGe:FORMat?
(see page 830)
<format> ::= {{BMP | BMP24bit} |
BMP8bit | PNG | NONE}

Table of Contents