244 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
6 Common (*) Commands
*RST (Reset)
(see page 1610)
Command Syntax
*RST
The *RST command places the instrument in a known state. This is the same as
pressing [Save/Recall] > Default/Erase > Factory Default on the front panel.
When you perform a factory default setup, there are no user settings that remain
unchanged. To perform the equivalent of the front panel's [Default Setup] key,
where some user settings (like preferences) remain unchanged, use the
:SYSTem:PRESet command.
Reset conditions are:
Acquire Menu
Mode Normal
Averaging Off
# Averages 8
Analog Channel Menu
Channel 1 On
Channel 2 Off
Volts/division 5.00 V
Offset 0.00
Coupling DC
Probe attenuation AutoProbe (if AutoProbe is connected),
otherwise 1.0:1
Vernier Off
Invert Off
BW limit Off
Impedance 1 M Ohm
Units Volts
Skew 0
Cursor Menu
Source Channel 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals