Root (:) Commands 7
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 279
Query Syntax :OPEE?
The :OPEE? query returns the current value contained in the Operation Status
Enable register as an integer number.
Return Format
<value><NL>
<value> ::= integer in NR1 format.
See Also • "Introduction to Root (:) Commands" on page 258
• ":AER (Arm Event Register)" on page 260
• ":CHANnel<n>:PROTection" on page 362
• ":OPERegister[:EVENt] (Operation Status Event Register)" on page 283
• ":OVLenable (Overload Event Enable Register)" on page 286
• ":OVLRegister (Overload Event Register)" on page 288
• "*STB (Read Status Byte)" on page 250
• "*SRE (Service Request Enable)" on page 248
• "Operation Status Event Register (:OPERegister[:EVENt])" on page 1571
• "Example: Checking for Armed Status" on page 1581
• "Example: Waiting for IO Operation Complete" on page 1586
10 --- --- (Not used.)
9 MTE Mask Test Event Event when mask test event occurs.
8-6 --- --- (Not used.)
5Wait
Trig
Wait Trig Event when the trigger is armed.
4 --- --- (Not used.)
3 Run Running Event when the oscilloscope is running (not
stopped).
2-0 --- --- (Not used.)
Table 89 Operation Status Enable Register (OPEE) (continued)
Bit Name Description When Set (1 = High = True), Enables:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals