Keysight InfiniiVision 4000 X-Series Oscilloscopes
Programmer's Guide
25 :MEASure Power Commands
These :MEASure commands are available when the power measurements and
analysis application is licensed and enabled.
Table 111 :MEASure Power Commands Summary
Command Query Options and Query Returns
:MEASure:ANGLe
[<source1>][,<source2
>] (see page 646)
:MEASure:ANGLe?
[<source1>][,<source2
>] (see page 646)
<source1>, <source2> ::=
{CHANnel<n>}
<n> ::= 1 to (# analog channels)
in NR1 format
<return_value> ::= the power
phase angle in degrees in NR3
format
:MEASure:APParent
[<source1>][,<source2
>] (see page 647)
:MEASure:APParent?
[<source1>][,<source2
>] (see page 647)
<source1>, <source2> ::=
{CHANnel<n>}
<n> ::= 1 to (# analog channels)
in NR1 format
<return_value> ::= the apparent
power value in NR3 format
:MEASure:CPLoss
[<source1>][,<source2
>] (see page 648)
:MEASure:CPLoss?
[<source1>][,<source2
>] (see page 648)
<source1>, <source2>
<source1> ::= {FUNCtion<m> |
MATH<m>}
<source2> ::= {CHANnel<n>}
<m> ::= 1 to (# math functions)
in NR1 format
<n> ::= 1 to (# analog channels)
in NR1 format
<return_value> ::= the switching
loss per cycle watts value in NR3
format