EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #643 background imageLoading...
Page #643 background image
:MEASure Power Commands 25
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 643
:MEASure:ONTime
[<source1>][,<source2
>] (see page 655)
:MEASure:ONTime?
[<source1>][,<source2
>] (see page 655)
<source1>, <source2> ::=
{CHANnel<n>}
<n> ::= 1 to (# analog channels)
in NR1 format
<return_value> ::= the time in
seconds in NR3 format
:MEASure:OPOWer (see
page 656)
:MEASure:OPOWer? (see
page 656)
<return_value> ::= the output
power value in NR3 format
:MEASure:PCURrent
[<source>] (see
page 657)
:MEASure:PCURrent?
[<source>] (see
page 657)
<source> ::= {CHANnel<n>|
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= the peak
current value in NR3 format
:MEASure:PLOSs
[<source>] (see
page 658)
:MEASure:PLOSs?
[<source>] (see
page 658)
<source> ::= {CHANnel<n>|
FUNCtion<m> | MATH<m> |
WMEMory<r>}
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1 to (# math functions)
in NR1 format
<r> ::= 1 to (# ref waveforms) in
NR1 format
<return_value> ::= the power loss
value in NR3 format
Table 111 :MEASure Power Commands Summary (continued)
Command Query Options and Query Returns

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals