Commands Quick Reference 5
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 203
Table 62 General :TRIGger Commands Summary
Command Query Options and Query Returns
:TRIGger:FORCe (see
page 1299)
n/a n/a
:TRIGger:HFReject {{0
| OFF} | {1 | ON}}
(see page 1300)
:TRIGger:HFReject?
(see page 1300)
{0 | 1}
:TRIGger:HOLDoff
<holdoff_time> (see
page 1301)
:TRIGger:HOLDoff?
(see page 1301)
<holdoff_time> ::= 60 ns to 10 s
in NR3 format
:TRIGger:HOLDoff:MAXi
mum <max_holdoff>
(see page 1302)
:TRIGger:HOLDoff:MAXi
mum? (see page 1302)
<max_holdoff> ::= maximum
holdoff time in seconds in NR3
format
:TRIGger:HOLDoff:MINi
mum <min_holdoff>
(see page 1303)
:TRIGger:HOLDoff:MINi
mum? (see page 1303)
<min_holdoff> ::= minimum
holdoff time in seconds in NR3
format
:TRIGger:HOLDoff:RAND
om {{0 | OFF} | {1 |
ON}} (see page 1304)
:TRIGger:HOLDoff:RAND
om? (see page 1304)
<setting> ::= {0 | 1}
:TRIGger:LEVel:ASETup
(see page 1305)
n/a n/a
:TRIGger:LEVel:HIGH
<level>, <source>
(see page 1306)
:TRIGger:LEVel:HIGH?
<source> (see
page 1306)
<level> ::= .75 x full-scale
voltage from center screen in NR3
format.
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger:LEVel:LOW
<level>, <source>
(see page 1307)
:TRIGger:LEVel:LOW?
<source> (see
page 1307)
<level> ::= .75 x full-scale
voltage from center screen in NR3
format.
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels)
in NR1 format
:TRIGger:MODE <mode>
(see page 1308)
:TRIGger:MODE? (see
page 1308)
<mode> ::= {EDGE | GLITch |
PATTern | TV | DELay | EBURst |
OR | RUNT | SHOLd | TRANsition |
SBUS{1 | 2}}
<return_value> ::= {<mode> |
<none>}
<none> ::= query returns "NONE"
if the :TIMebase:MODE is ROLL or
XY