Root (:) Commands 7
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 257
:OPEE <n> (see
page 278)
:OPEE? (see page 279) <n> ::= 15-bit integer in NR1
format
n/a :OPERegister:CONDitio
n? (see page 280)
<n> ::= 15-bit integer in NR1
format
n/a :OPERegister[:EVENt]?
(see page 283)
<n> ::= 15-bit integer in NR1
format
:OVLenable <mask>
(see page 286)
:OVLenable? (see
page 287)
<mask> ::= 16-bit integer in NR1
format as shown:
Bit Weight Input
--- ------ ----------
10 1024 Ext Trigger Fault
9 512 Channel 4 Fault
8 256 Channel 3 Fault
7 128 Channel 2 Fault
6 64 Channel 1 Fault
4 16 Ext Trigger OVL
3 8 Channel 4 OVL
2 4 Channel 3 OVL
1 2 Channel 2 OVL
0 1 Channel 1 OVL
n/a :OVLRegister? (see
page 288)
<value> ::= integer in NR1
format. See OVLenable for <value>
:PRINt [<options>]
(see page 290)
n/a <options> ::= [<print
option>][,..,<print option>]
<print option> ::= {COLor |
GRAYscale | PRINter0 | BMP8bit |
BMP | PNG | NOFactors | FACTors}
<print option> can be repeated up
to 5 times.
:RUN (see page 291) n/a n/a
n/a :SERial (see page 292) <return value> ::= unquoted
string containing serial number
:SINGle (see page 293) n/a n/a
Table 83 Root (:) Commands Summary (continued)
Command Query Options and Query Returns