496 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
20 :FUNCtion<m> Commands
:FUNCtion<m>:SCALe
(see page 1610)
Command Syntax
:FUNCtion<m>:SCALe <scale value>[<suffix>]
<m> ::= 1 to (# math functions) in NR1 format
<scale value> ::= vertical units/div value in NR3 format
<suffix> ::= {V | dB}
The :FUNCtion<m>:SCALe command sets the vertical scale, or units per division,
of the selected function. Legal values for the scale depend on the selected
function.
Query Syntax
:FUNCtion<m>:SCALe?
The :FUNCtion<m>:SCALe? query returns the current scale value for the selected
function.
Return Format
<scale value><NL>
<scale value> ::= vertical units/div value in NR3 format
See Also • "Introduction to :FUNCtion<m> Commands" on page 460
• ":FUNCtion<m>:DISPlay" on page 469
• ":FUNCtion<m>:OPERation" on page 490
• ":FUNCtion<m>:RANGe" on page 494
NOTE
Automatic vertical scaling of the math function waveform occurs when the function display is
turned from off to on (see :FUNCtion<m>:DISPlay) or, if the function is already on, when the
operation is changed (see :FUNCtion<m>:OPERation). If you want to change the math
function's vertical scaling, you should do it after the math function display is turned on or after
the operation is changed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals