EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
494 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
20 :FUNCtion<m> Commands
:FUNCtion<m>:RANGe
(see page 1610)
Command Syntax
:FUNCtion<m>:RANGe <range>
<m> ::= 1 to (# math functions) in NR1 format
<range> ::= the full-scale vertical axis value in NR3 format.
The :FUNCtion<m>:RANGe command defines the full-scale vertical axis for the
selected function.
Query Syntax
:FUNCtion<m>:RANGe?
The :FUNCtion<m>:RANGe? query returns the current full-scale range value for
the selected function.
Return Format
<range><NL>
<range> ::= the full-scale vertical axis value in NR3 format.
See Also "Introduction to :FUNCtion<m> Commands" on page 460
":FUNCtion<m>:SCALe" on page 496

Table of Contents

Related product manuals