EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1560 background imageLoading...
Page #1560 background image
1560 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
41 Status Reporting
The status register bits are described in more detail in the following tables:
• Table 82
• Table 80
• Table 90
• Table 91
• Table 93
• Table 85
• Table 86
• Table 88



Run bit set if oscilloscope not stopped
Wait
Trig
Run
Arm
Reg
AER ?
:OPERegister:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable (Mask) Register
10234567
OPER TRGMAVESB
RQS/
MSS
TRG
Reg
TER?
Trigger Event Register
*SRE
*SRE?
Service Request Enable (Mask) Register
*STB?
Status Byte Register
Outp ut
Queue
SRQ
*ESR?
(Standard) Event Status Register
*ESE
*ESE?
(Standard) Event Status Enable (Mask) Register
PON URQ EXE DDE QYE RQL OPCCM E
OVLR
USR
MSG
OR
10234567
OR
1023456798101112131415
Se rvice Requ est
OR
Wait
Trig
Run
:OPERegister[:EVENt]?
Operation Status Event Register
OVLR
11 5 3
From Overload
Event Registers
From Mask Test
Event Registers
MTE
MTE
9
IOF
IOC
IOF
IOC
1314
From Hardware
Event Registers
12
HWE
HWE
RUI
Enab
RUI
Enab
Set if remote user interface is enabled

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals