Width
For the ranges "Shorter"
and "Longer", the width defines the maximum and minimum
pulse width, respectively.
For the ranges "Within" and "Outside", the width defines the center of a range which is
defined by the limits "±Delta".
Remote command:
TRIGger<m>:PARallel:WIDTh:WIDTh on page 2434
±Delta
Defines a range around the given width value.
The combination "Range" = Within and "±Delta" = 0 triggers on pulses with a pulse
width that equals "Width".
The combination "Range" = Outside and "±Delta" = 0 means to trigger on pulse widths
"Width".
Remote command:
TRIGger<m>:PARallel:WIDTh:DELTa on page 2435
14.3.1.4 Timeout
The timeout trigger event checks if the trigger source signal stays above or below the
threshold voltage for a specified time lapse. In other words, the event occurs if the
state condition remains unchanged for the specified time.
Y
ou can use the timeout trigger on a single digital channel, or a logical combination of
digital channels.
Figure 14-3: Timeout trigger settings for trigger source = logical combination of digital channels
Range
Sets the state condition:
Trigger