Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1143User Manual 1801.6687.02 ─ 05
To measure and display the inrush current, the instrument uses the following measure-
ments and waveforms:
"MG 1" to "MG 5" to measure the inrush current of "Gate 1" to "Gate 5"
The used resources are listed in the "Details" tab. See also: Chapter 19.2.3, "Details
tab", on page 1135.
The following remote commands are used for handling the measurement results:
POWer:INRush:GATE<m>:VALue on page 2530
POWer:INRush:REPort:ADD on page 2531
19.3.2.2 Configuring inrush current
For details of the configuration settings, see Chapter 19.3.2, "Inrush current",
on page 1142.
1. Select "Analysis" > "Power".
2. Under "Input", select "Inrush Current".
3. Connect the current probe to the oscilloscope.
4. Select the correct channel for the "Current Source".
5. Select "Vertical" > "Probe Setup" > "Channel" and set your probe parameters.
6. Connect the probes to the DUT as shown in the "Channels" tab:
Input