Acquisition and waveform setup
R&S
®
RTO6
156User Manual 1801.6687.02 ─ 05
AC Coupling
Enables AC coupling in the R&S R
T
‑ZPR power rail probes, which removes DC and
very low-frequency components. The R&S RT‑ZPR probe requires 50 Ω input termina-
tion, for which the channel AC coupling is not available. The probe setting allows AC
coupling also at 50 Ω inputs.
Remote command:
PROBe<m>:SETup:ACCoupling on page 1353
Copy to offset
Sets the measured ProbeMeter value as offset. Thus, the value is considered in mea-
surements.
Remote command:
PROBe<m>:SETup:ADVanced:PMToffset on page 1353
5.6.4.4 R&S RT-ZHD high-voltage differential probes
Figure 5-12: Probe setup for R&S
RT-ZHD probes
The offset is the differential offset. See "Of
fset
" on page 139.
Specific settings for R&S RT-ZHD probes are the following:
Probes