Message
Sets the message to be triggered or searched for
.
The setting is used by symbolic trigger and symbolic search. Thus, you always trigger
on and search for the same message if symbolic trigger and symbolic search is used at
the same time.
Remote command:
TRIGger<m>:
CAN:SYMBolic:MSGValue on page 1832
Signal trigger
Enables the trigger on a specific signal value that is part of the selected message.
Remote command:
TRIGger<m>:CAN:SYMBolic:TSIGnals on page 1832
Signal
Sets the signal name to be triggered or searched for.
The setting is used by symbolic trigger and symbolic search. Thus, you always trigger
on and search for the same signal if symbolic trigger and symbolic search is used at
the same time.
Remote command:
TRIGger<m>:CAN:SYMBolic:SIGValue on page 1832
Condition
Sets the operator to set a specific data pattern or symbolic value ("Equal" or "Not
equal") or a data range.
Remote command:
TRIGger<m>:CAN:DCONdition on page 1804
Value, Value min
Defines the data pattern or selects a symbolic data value.
Remote command:
TRIGger<m>:CAN:SYMBolic:DMIN on page 1833
TRIGger<m>:CAN:SYMBolic:SGEValue on page 1833
Value max
The second data pattern is required to specify a range with conditions "In range" and
"Out of range".
Remote command:
TRIGger<m>:CAN:SYMBolic:DMAX on page 1833
13.7.6.2 Symbolic decode waveform
If a DBC file is applied, the symbolic names from the file are applied to the display of
the decoded data. The result table lists the signal values and units in the "Symbolic
Data" column, and the comb display shows the signal names in addition to the signal
values and units.
CAN and CAN FD (option R&S R