Measurements
R&S
®
RTO6
335User Manual 1801.6687.02 ─ 05
8.2.7.1 Overview of spectrum measurements
Table 8-7: Spectrum measurements
Meas. type Description, result
Channel power Power integrated over the sample values defined by a center frequency
and a bandwidth; based on a defined impedance. The result is given in
dBm.
To get best results, enable "Measure all events in each acquisition
"
on page 362, and set the "Limit" to the maximum value. In particular,
these settings are important when measuring pulsed signals.
%
Occupied bandwidth From the defined center frequency, symmetric sample value pairs to the
left and right are integrated until a user-defined percentage of the total
power is reached. The occupied bandwidth is the difference between
the frequencies at which the requested power was reached.
N
Bandwidth n dB down bandwidth; the samples to the left and right of the peak
value are analyzed until the n dB threshold is exceeded. The frequen-
cies at which the threshold is exceeded define the limits of the reques-
ted bandwidth.
THD[dB], THD[%]
Total harmonic distor-
tion
Power sum of the harmonic waves divided by the power of the funda-
mental wave:
1
2
P
P
THD
n
n
Automatic measurements