Mixed signal option (MSO, R&S RTO6-B1)
R&S
®
RT
O6
997User Manual 1801.6687.02 ─ 05
Table 14-1: Trigger types and digital trigger sources
Trigger type Trigger source is
Digital channel Logic combination of
digital channels
Parallel bus
Edge X X
Width X X
Timeout X X
Setup & Hold X
X
State
X X
Pattern (with holdoff)
X X
Serial Pattern X X
For details, see: Chapter 14.3.1, "Trigger settings for digital signals and parallel buses",
on page
997.
Also, you can define trigger holdoff conditions. See Chapter 6.4, "Mode / Holdoff",
on page 220.
14.3.1 Trigger settings for digital signals and parallel buses
Depending on the selected source, the instrument provides the appropriate trigger
types and the corresponding trigger settings.
The settings in the "Setup" tab are:
● Basic trigger settings.............................................................................................997
● Edge
......................................................................................................................998
● Width.....................................................................................................................999
● Timeout............................................................................................................... 1001
● Setup & Hold.......................................................................................................1002
● State....................................................................................................................1004
● Pattern.................................................................................................................1005
● Serial pattern.......................................................................................................1007
14.3.1.1 Basic trigger settings
The basic trigger settings for MSO are the trigger source and the trigger type. They are
selected in the upper part of the "Trigger" dialog box.
Make sure that the trigger sequence is set to "A only".
Also, you can define trigger holdoff conditions. See Chapter 6.4, "Mode / Holdof
f",
on page
220.
Trigger