Remote control commands
R&S
®
RTO6
1403User Manual 1801.6687.02 ─ 05
Parameters:
<WidthDelta> Range: 0 to 10
Increment: 100E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
24.9.9 Slew rate trigger
TRIGger<m>:SLEW:
SLOPe..........................................................................................1403
TRIGger<m>:LEVel<n>:SLEW:LOWer........................................................................... 1403
TRIGger<m>:LEVel<n>:SLEW:UPPer............................................................................1403
TRIGger<m>:SLEW:RANGe......................................................................................... 1404
TRIGger<m>:SLEW:RATE............................................................................................1404
TRIGger<m>:SLEW:DELTa...........................................................................................1405
TRIGger<m>:SLEW:SLOPe <Slope>
Selects the edge type for the trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger
, 2 = B-trigger
, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative | EITHer
See Chapter 24.4.3, "Slope parameter", on page 1285.
*RST
: POSitive
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:SLEW:LOWer <Level>
TRIGger<m>:LEVel<n>:
SLEW:UPPer <Level>
Set the lower and upper voltage thresholds. When the signal crosses a level, the slew
rate measurement starts or stops depending on the selected slope.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger
, 3 = R-trigger
<n> 1..31
Indicates the trigger source:
1...4 = channel 1...4
5...27 = not available
28...31 = differential signals Dif
f1, Diff2 and Common1, Com-
mon2
Trigger