Remote control commands
R&S
®
RTO6
2519User Manual 1801.6687.02 ─ 05
Parameters:
<CdrInUse> ON | OFF
*RST: OFF
Usage: Asynchronous command
CDR:HARDware:SAMTime <UntIntvlOffs>
Sets a sampling time for the clock signal, an of
fset for the clock edge in relation to the
bit start. The clock edge sets the beginning of the unit interval. The sampling time is a
number between 0 and 1. Value 0 sets the clock edge to the beginning of the bit
period; value 0.5 sets the clock edge to the middle of the bit period.
Parameters:
<UntIntvlOffs> Range: 0 to 1
Increment: 0.01
*RST: 0.5
Default unit: UI
Usage: Asynchronous command
24.22.6 Eye mask testing
The eye mask testing is available if the option R&S R
T
O6-K12 is activated on the
instrument.
The commands for eye mask testing are described in Chapter 24.14.6, "Eye mask test-
ing", on page 1599.
24.23 Power analysis (option R&S RTO6-K31)
● General............................................................................................................... 2520
● Deskew............................................................................................................... 2520
● Report................................................................................................................. 2522
● Power quality.......................................................................................................2527
● Inrush current......................................................................................................2529
● Current harmonic................................................................................................ 2531
● Modulation analysis.............................................................................................2533
● Dynamic ON resistance...................................................................................... 2535
● Slew rate............................................................................................................. 2536
● S.O.A...................................................................................................................2539
● Turn on/off...........................................................................................................2541
● Switching loss..................................................................................................... 2543
● Power efficiency..................................................................................................2546
● Ripple..................................................................................................................2547
● Transient response............................................................................................. 2554
● Spectrum.............................................................................................................2557
Power analysis (option R&S R
T
O6-K31)