Power analysis (option R&S RTO6-K31)
R&S
®
RT
O6
1160User Manual 1801.6687.02 ─ 05
Current probe
19.4.3.1 Slew rate results
The results of "Slew Rate"
measurements are provided in two ways:
The diagram shows the graphical presentation of:
the voltage waveform
the current waveform
a waveform of the derivative of voltage and current
The result box displays the numeric measurement results. For a detailed descrip-
tion, see Table 19-2.
To measure and display the slew rate, the instrument uses the following measurements
and waveforms:
"MG 8" to measure the amplitude of the current or voltage waveform
"M2" Math 4 to calculate the time derivative of the current or voltage waveform
"Cu 1" Cursor 1 to determine the measurement area
The used resources are listed in the
"Details" tab. See also: Chapter 19.2.3, "Details
tab", on page 1135.
The following remote commands are used for handling the measurement results:
POWer:SLEWrate:RESult:ACTual? on page 2538
POWer:SLEWrate:RESult:AVG? on page 2538
POWer:SLEWrate:RESult:EVTCount? on page 2538
POWer:SLEWrate:RESult:NPEak? on page 2538
POWer:SLEWrate:RESult:PPEak? on page 2538
POWer:SLEWrate:RESult:RMS? on page 2538
POWer:SLEWrate:RESult:STDDev? on page 2539
POWer:SLEWrate:RESult:WFMCount? on page 2539
POWer:SLEWrate:REPort:ADD on page 2538
19.4.3.2 Configuring slew rate
For details of the configuration settings, see Chapter 19.4.3.3, "Slew rate settings",
on page 1162.
1. Select "Analysis" > "Power".
Control loop