Deembedding (option R&S RTO6-K121)
R&S
®
RT
O6
1205User Manual 1801.6687.02 ─ 05
Remote command:
DEEMbedding<m>:
COMPonent<n>:
LOAD:IMPedance? on page 2565
PROBe<m>:DEEMbedding:LOAD:IMPedance on page 1361
20.2.4 R&S RT‑ZM probe settings
If an R&S RT‑ZM modular probe is connected to a channel input, the probe is detected
by the oscilloscope and automatically added to the measurement path. The S-parame-
ters of the probe, the tip modules, and the probe group delay are known to the oscillo-
scope and used for deembedding.
Probe settings that are important for deembedding can be set directly in the deembed-
ding dialog.
To get all probe deembedding settings:
1. In the "Deembedding" > "Components" > "RT-ZMxx" tab, select "More Settings".
2. In the "Probes Setup" dialog box, select "Probe Deembedding".
The "Probe Deembedding" dialog box is described in Chapter 5.6.5.2, "Deembedding
for modular probes", on page 159.
For details on probe parameters, see:
●
"Select probe tip module, Pin spacing" on page 159
●
"Probe mode" on page 158
●
"DM Offset, CM Offset, P Offset, N Offset" on page 159
20.2.5 Input settings for other probes
The "Input" component is used for all probes except for R&S RT‑ZM, R&S RT-ZD and
R&S RT-ZS probes with Rohde & Schwarz interface, and for cables or other compo-
nents that are used to connect the measurement setup to the oscilloscope. For these
input components, the horizontal skew and vertical settings are considered during
deembedding. You can set the skew directly in the "Deembedding" > "Components" >
"Input" tab, or in "Horizontal" menu > "Skew". The button "More Settings" opens the
"Channels" settings.
The input component is also used for deembedded differential signals, see Chap-
ter 5.8.3, "Deembedding of differential signals", on page 184.
Deembedding settings