Instrument setup
R&S
®
RTO6
98User Manual 1801.6687.02 ─ 05
2. Select the "Touchscreen" tab.
3.
T
ap "Calibrate".
A blinking cross appears in the lower left corner of the screen.
4. Touch and hold the cross until "OK" is shown.
5. Repeat this action for the crosses in the other corners.
6. Tap the Rohde & Schwarz logo to display the instrument's user interface.
4.6 Preset setup
A user-defined preset contains the complete instrument setup including display set-
tings, except for transparency and intensity
. Y
ou can save the current configuration to a
preset file, and load a previously saved preset file. You can then specify that these set-
tings are to be applied with the [Preset] key.
4.6.1 User-defined preset
Access: "Menu" > "Settings" > "Preset" > "Settings" tab
Preset
Performs a user-defined preset of the instrument.
Saveset to be loaded on preset
The file name with extension .dfl to load or to save the settings to.
Remote command:
USRDefined:
RST:NAME on page 1321
USRDefined:RST:SAVE on page 1322
USRDefined:RST:OPEN on page 1321