Remote control commands
R&S
®
RTO6
1263User Manual 1801.6687.02 ─ 05
REFLevel2:LDETection MANual
REFLevel2:LMODe ABS
REFLevel2:USRLevel UREF
REFLevel2:ABSolute:LLEVel 0
REFLevel2:ABSolute:BDIStance 0.02
REFLevel2:ABSolute:ULEVel 0.2
REFLevel2:ABSolute:TDIStance 0.03
*OPC?
REFLevel2:ABSolute:MLEVel?
<-- 0.1
Automatic level detection, peak probability
Reference levels are set to the signal levels with the highest probability values for
waveform C3W1 (= suf
fix 8).
REFLevel8:LDETection Auto
REFLevel8:AUTO:MODE PPRobability
*OPC?
24.3.2.3
Waveform histograms
In the following example *OPC prevents overlapping execution of asynchronous com-
mands.
Creating and reading histograms........................................................................1263
Exporting histogram data to file...........................................................................1264
Transferring histogram data................................................................................1264
Creating and reading histograms
The example creates a histogram, activates two measurements (mean and standard
deviation measurements of Histogram1), and queries the results of both measure-
ments.
Command description in:
Chapter 24.12.1, "General settings", on page 1480
Chapter 24.12.8.2, "Histogram measurement", on page 1527
Chapter 24.12.2, "Results", on page 1485
LAY:HIST:ADD 'Histogram1', C1W1, -2.5E-007, 2.5E-007, -1.32, 5.35, OFF, VERT *OPC?
MEAS1 ON
MEAS1:HIST:SEL 'Histogram1'
MEAS1:CAT HIST
MEAS1:MAIN HME
*OPC?
MEAS2 ON
MEAS2:HIST:SEL 'Histogram1'
MEAS2:CAT HIST
MEAS2:MAIN HSTD
Programming examples

Table of Contents

Other manuals for R&S RTO6

Related product manuals