Remote control commands
R&S
®
RTO6
2451User Manual 1801.6687.02 ─ 05
Parameters:
<Depth> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
WGENerator<m>:MODulation:PWM:FREQuency <Frequency>
Sets the frequency of the modulation waveform for pulse width modulation, if
WGENerator<m>:MODulation:
TYPE is set to PWM.
Suffix:
<m>
.
1..2
Parameters:
<Frequency> Range: 1E-3 to 1E+6
Increment: 1
*RST: 1000
Default unit: Hz
WGENerator<m>:MODulation:PWM:SYMMetry <RampSymmetry>
Sets the symmetry, the percentage of time the ramp modulation waveform is rising, for
pulse width modulation, if WGENerator<m>:
MODulation:TYPE is set to PWM.
Suffix:
<m>
.
1..2
Parameters:
<RampSymmetry> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
WGENerator<m>:MODulation:PWM[:FUNCtion] <SignalType>
Sets the type of the modulation signal for the pulse width modulation, if
WGENerator<m>:
MODulation:TYPE is set to PWM.
Suffix:
<m>
.
1..2
Parameters:
<SignalType> SINusoid | SQUare | RAMP
*RST: SINusoid
24.19.1.4 Sweep
WGENerator<m>:SWEep:FSTart
...................................................................................2452
WGENerator<m>:SWEep:TIME.....................................................................................2452
WGENerator<m>:SWEep[:FEND]..................................................................................2452
Waveform generator (option R&S R
T