Protocol analysis
R&S
®
RTO6
924User Manual 1801.6687.02 ─ 05
Threshold High
Sets the threshold high of the signal, if
"Signal type"
> "Differential".
Remote command:
BUS<m>:EBTB:DIFFerential:THRHigh on page 2332
Threshold Low
Sets the threshold low of the signal, if "Signal type" > "Differential".
Remote command:
BUS<m>:EBTB:DIFFerential:THRLow on page 2333
Threshold D+
Sets the high threshold (D+) of the signal, if "Signal type" > "Single ended".
Remote command:
BUS<m>:EBTB:DPLus:THReshold on page 2334
Threshold D-
Sets the high threshold (D-) of the signal, if "Signal type" > "Single ended".
Remote command:
BUS<m>:EBTB:DMINus:THReshold on page 2333
Find thresholds
Executes the measurement of reference levels and sets the thresholds to the middle
reference level of the measured amplitude.
Remote command:
BUS<m>:SETReflevels on page 1671
Bitrate
Sets the number of transmitted bits per second.
Remote command:
BUS<m>:EBTB:BITRate on page 2332
Estimate Bitrate
Starts a software algorithm for the automatic determination of the bitrate. The value is
displayed in the "Estimated Bitrate".
Remote command:
BUS<m>:EBTB:BITDetermi on page 2336
BUS<m>:EBTB:EBTRate? on page 2332
Show thresholds
If enabled, the threshold levels are displayed in the diagram.
Remote command:
BUS<m>:THReshold on page 1671