Acquisition and waveform setup
R&S
®
RTO6
146User Manual 1801.6687.02 ─ 05
● Probe attributes.....................................................................................................
168
● Calibration results................................................................................................. 169
● Probe adapter R&S R
T‑Z2T..................................................................................169
5.6.1 Shared probe settings
Some of the settings in the "Probes" tab are available for all probes.
An external attenuation can be set on the "Other" tab, see "External attenuation: Scale,
Attenuation"
on page 142.
The shared probe settings are:
Type, Name, Bandwidth..............................................................................................146
Detect AutoZero, Use AutoZero..................................................................................147
Type, Name, Bandwidth
The fields show the characteristics of a recognized or predefined probe for information.
If the instrument cannot recognize the probe, and the probe is not known, the "Type" is
"None", and the other fields are empty.
Remote command:
PROBe<m>:SETup:TYPE? on page 1344
PROBe<m>:SETup:NAME? on page 1344
PROBe<m>:SETup:BANDwidth? on page 1345
Probes