I/Q software interface (Option R&S RTO6-K1
1)
R&S
®
RTO6
1064User Manual 1801.6687.02 ─ 05
Sideband (complex IF input)
Defines the sideband and the frequency position of complex modulated input signal in
IF range.
"Normal/Upper"
: Upper sideband in normal position
f
c
Input signal: normal
position, upper sideband
f
c
x(t)e
j2
π
f
c
t
-
Down-conversion and
filtering
"Normal/Lower": Lower sideband in normal position
- f
c
- f
c
x(t)e
j2
π
f
c
t
Down-conversion, filtering,
and complex conjugation
Input signal: normal
position,
lower sideband
"Inverse/Upper"
: Upper sideband in inverse position
f
c
Input signal: inverse
position, upper sideband
f
c
x(t)e
j2
π
f
c
t
-
Down-conversion, filtering,
and complex conjugation
"Inverse/Lower": Lower sideband in inverse position
- f
c
- f
c
x(t)e
j2
π
f
c
t
Input signal: inverse
position, lower sideband
Down-conversion and
filtering
Remote command:
CHANnel<m>:IQ:SBIF on page 2476
17.3 I/Q data output