Triggers
R&S
®
RTO6
209User Manual 1801.6687.02 ─ 05
Use data level for clock level
Sets the trigger levels and hysteresis values for all channels to the values of the cur-
rently selected trigger source. The function af
fects only the levels defined for the
selected condition. The hysteresis of the external trigger input is an independent value,
and it is not af
fected by level coupling.
Remote command:
TRIGger<m>:SCOupling on page 1407
Setup time
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
The setup time can be negative. In this case, the hold time is always positive. If you set
a negative setup time, the hold time is adjusted by the instrument.
Remote command:
TRIGger<m>:SETHold:STIMe on page 1407
Hold time
Sets the minimum time after the clock edge while the data signal must stay steady
above or below the data level.
The hold time can be negative. In this case, the setup time is always positive. If you set
a negative hold time, the setup time is adjusted by the instrument.
Remote command:
TRIGger<m>:SETHold:HTIMe on page 1406
6.3.11 State
The state trigger is a qualified edge trigger. It combines the edge trigger settings with
trigger qualification.
Trigger types