EasyManua.ls Logo

R&S RTO6 - 8.2.13.1 Performing limit checks; 8.2.13.2 Limit and margin settings

R&S RTO6
2677 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements
R&S
®
RTO6
364User Manual 1801.6687.02 ─ 05
Statistical results. Limit and margin violations of statistical results are indicated by
icons in the result box. These violations do not initiate an action.
See also: Chapter 8.2.2.1, "Measurement status", on page 302
8.2.13.1 Performing limit checks
1. Open the menu and select "Measure".
2.
Select the "Advanced"
tab.
3. Select the subtab of the measurement group that you want to configure.
4. Under "Type", select "Limit only" to distinguish only between valid and non-valid
values.
Select "Margin&Limit" to check two values, where the margin is still valid, while the
limit is not.
5. Tap "Setup". Define the valid value range for each active measurement.
The margins must always be within the valid value range. If necessary, the limit or
margin values are adapted to match the selected valid range.
See also "Upper limit, Lower limit, Upper margin, Lower margin, Valid range"
on page 365
6. Define what happens when the defined limits and margins are exceeded.
For each action, define when the instrument starts it:
If the limits or margins are exceeded
If the measurement is completed without limit violations
Not at all
As a result of the limit check, the specified actions are performed and an icon indi-
cates the status in the result box.
8.2.13.2 Limit and margin settings
Access: [Measure] > "Advanced" tab
If a check is selected, tap "Setup" to set the limit and margin values and the range of
valid measurement results.
Automatic measurements

Table of Contents

Other manuals for R&S RTO6

Related product manuals